GSI TS260 Stop Trim Interface Card

GSI 229.260.00 – TS260 Stop Trim Board [LL Server]

gsi laser

The TS260 Stop Trim Interface Card (slot B18) is typically used in conjunction with the TS900 Test System and/or the HP 3458A Multimeter. The TS260 is a multichannel handshake signal interface card that conditions the signals that
pass between an M310 and the external test instruments. These instruments are typically selected to support resistor trim operations. The Stop Trim card gets power and ground from the backplane. Other signals are exchanged through two 20-pin header connectors, four BNC connectors, and a 40-pin header connector.
There have been variations of the TS260 used in several M310 installations. The following material applies to those variants only in a general sense.
Operating Modes – There are two operating modes for the TS260: Test Mode
and Trim Mode. This is controlled by the bit status on J3-29 (B1):

B1 = 1 Trim Mode
B1 = 0 Test Mode

Connections – The actual cable connections will vary based upon the system
configuration. The most commonly used include:

• J3 is normally connected to the TS227 board in slot A4.

• J4 (SRQ EXT OUT) is normally connected to the EXT SIG OUT connection on
the HP 3458A or its equivalent. This signal indicates that the measured values has
reached the predefined setpoint for that measurement and that trimming should be stopped.

• J5 (MUX CH CL/AUX 1) is used to detect that the Multiplexer channel has been
successfully changed. This is not normally used in most M310 systems.

• J6 (MUX INC/AUX I/O) is used to increment the HP Multiplexer (or equivalent)
that is used to step between test probes on the probe card. This is not normally
used in most M310 systems.

• J7 (EXT TRIG) is connected to the EXT TRIG IN connection of the HP 3458A or
its equivalent. This signals the meter to make a measurement

In order to minimize the time required to perform maintenance or repairs on your system, it is advisable to maintain a local supply of spare parts.

The quantity and nature of these spare parts is determined by the Periodic Maintenance requirements of the system, and by the number of systems that are installed at your location.

It may also be important to consider the difficulty of shipping components to your location. Spare parts can be ordered either through Specialized Electronic Services. Toll free phone numbers are available. Call Now!

You can also email inquiries regarding the availability of spare parts can be addressed to:

Note: When ordering parts such as galvanometers, lasers or expensive optics, consult SES before you consult other vendors. SES has years of experience and will give you the best pricing and service in the industry.